Advanced Calibration Designs Inc QC 100 Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Telephone 520 290 2855 Fax 520 290 2860 Toll Free 800 737 0223 Email acd goACD com Web www goACD com P N 910 1100 00 Calibration Gas Bump Tester Instruction Manual
Instruction Manual QC 100 QuickCheck System TABLE OF CONTENTS I QC 100 Operating Instructions Page 2 II Accessory Items Parts List Page 4 III Specifications Page 5 910 1050 00
I QC 100 OPERATING INSTRUCTIONS 1 Remove and discard any used QuickCheck Vial QCV If a vial is left attached to the QC 100 during storage the flow path may become saturated This is especially a concern if more than one gas type of QCV is being used 2 Press the On Off button to turn on the QC 100 A green light indicates normal operation A red light indicates a low battery condition If a red light occurs replace the two AA batteries with fresh alkaline batteries and repeat step two Note Do not use a metal object to remove the batteries as this can short out the electrical circuit 3 Select the QCV with the gas type desired 4 Squeeze the QCV firmly at the point where the label reads Pinch here to break inner ampoule This allows the liquid to absorb into the wicking material and start releasing the test gas 5 Install the QCV into the QC 100 To install insert the QCV snuggly into the opening on the right side of the instrument near the top 6 Attach any tubing or calibration adapters if needed Note A three foot section of chemically resistant tubing is included with the QC 100 If inappropriate tubing or adapters are used the chemical may be lost in the sample system without reaching the sensor to be tested 7 Place the end of the QC 100 tubing or calibration adapter at the sensor to be tested 8 If an appropriate response is not achieved the output of gas can be varied by opening turning counter clockwise or closing turning clockwise the needle valve on the top side of the QC 100 9 Repeat steps 7 and 8 for any other sensors to be tested Note Do not continue testing if the QCV is depleted Page 2
10 Turn off the QC 100 by pressing the On Off switch Remove and discard the used QCV 11 Note Once the inner ampoule is broken the QCV should be quickly inserted into the QC 100 Opened QCV deplete quickly without being installed Once testing is finished please remove and discard the QCV before storing the QC 100 Opened QCV should not be stored for longer than 24 hours Page 3
II ACCESSORY ITEMS PARTS LIST The following items and spare parts are available for the QC 50 P N Description 525 0001 12 NH3 Ammonia 12 Pack 525 0002 12 HCl Hydrogen Chloride 12 Pack 525 0003 12 C7H8 Toluene 12 Pack 525 0004 12 C6H14 Hexane 12 Pack 715 0405 0X Hose with Quick Connect 5 730 0415 00 Hard Body Carrying Case 730 1050 00 Nylon Carrying Case Page 4
III SPECIFICATIONS Useful Vial Life Up to 8 hours per vial Warranty One year Dimensions QC 100 6 0 x 2 3 x 1 2 15 2 x 5 8 x 3 0 Dimensions QCV 0 4 x 1 3 1 1 x 3 4cm Weight QC 100 8oz 220gm Weight QCV 0 14oz 4gm Operating Temperature 5 50 C Relative Humidity intermittent use 0 100 Battery Power 2 Alkaline AA Battery Life 10 hours Page 5
Manufactured by Advanced Calibration Designs Inc 2024 W McMillan Street Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 email ACD goacd com website www GoACD com
Advanced Calibration Designs Inc QC 100 Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Telephone 520 290 2855 Fax 520 290 2860 Toll Free 800 737 0223 Email acd goACD com Web www goACD com P N 910 1100 00 Calibration Gas Bump Tester Instruction Manual