Return to flip book view

MiniBump Instruction Manual

Page 1

Advanced Calibration Designs Inc MiniBump Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Telephone 520 290 2855 Fax 520 290 2860 Toll Free 800 737 0223 Email acd goACD com Web www goACD com P N 910 0550 00 Rel IV Calibration Gas Bump Tester Instruction Manual Rel IV

Page 2

WARNING This instrument generates test gas for toxic gas detectors The instruction manual should be read and understood prior to operation of the instrument Failure to operate the instrument correctly can lead to improper testing of your detection system This instrument conforms to the protection requirements of the EC DIRECTIVE 2014 30 EU on Electromagnetic Compatibility EMC in accordance with the provisions of Statutory Instrument 2372 The following standards have been applied EN 61000 6 1 2007 10 EN 61326 1997 A1 1998 A22001 A3 2003 Class A EN61326 1 2013 1

Page 3

2

Page 4

Instruction Manual MiniBump Calibration Gas Instrument Release IV P N 910 0550 00 Table of Contents WARNING 1 Table of Contents 3 I GENERAL DESCRIPTION 5 Internal Micro Pump 5 Electrochemical Generating Source 5 Alkaline A Batteries 5 Microprocessor Based Circuitry 6 II START UP MINIBUMP SOURCE 7 III NORMAL OPERATION MINIBUMP SOURCE 8 IV START UP MINICAL SOURCE 9 V NORMAL OPERATION MINICAL SOURCE 10 VI NON NORMAL OPERATION 11 VII CELL LIFE 12 VIII MAINTENANCE 13 ON OFF Switch 13 Caution 14 IX ACCESSORY ITEMS PARTS LIST 15 X STANDARD WARRANTY 16 3

Page 5

4

Page 6

I GENERAL DESCRIPTION The MiniBump is a small battery powered portable electrochemical gas generator designed to functionally test and or calibrate toxic gas sensors and alarm systems The MiniBump is designed to bump a gas detection system into alarm within approximately 30 seconds when used with a MiniBump generating source When used with a MiniCal generating source the unit will provide calibration gas of the type and concentration determined by the generating source to calibrate gas detection systems The MiniBump uses the following components to produce the calibration gas air mixture Internal Micro Pump A small air pump draws in ambient air to blend with the electrochemically generated gas The Release IV MiniBump instruments are designed to operate at 5 LPM liters per minute Electrochemical Generating Source The electrochemical generating source contains an electrolyte solution and either inert or consumable electrodes depending upon the gas being generated Gas is generated in the solution and is delivered out of the source by the internal pump to the detection system Alkaline A Batteries A set of two fully charged alkaline AA batteries provides approximately 10 12 hours of operation To replace the batteries open the battery cover with either a fingernail or small flat object It is important that when changing the batteries in the instrument to not use a metal object such as a screw driver Using a metal object to remove the batteries will frequently blow the pico fuse in the unit preventing the unit from powering up 5

Page 7

NOTE Rechargeable batteries will provide substantially fewer hours of operation and will self discharge while the instrument is in storage Users should also be aware that use of rechargeable batteries will void the CSA approval of the MiniBump for intrinsic safety Microprocessor Based Circuitry When powered up the microprocessor checks for what type of generating source is inserted the gas type and concentration as well as if the source is a bump type source and the remaining source life It also monitors battery life and indicates when either a low battery or exhausted source condition occurs The circuit board is attached directly to the bottom case via solder joints to the battery terminals If damaged or inoperable it should be replaced as part of the case rather than removed from these terminals 6

Page 8

II START UP MINIBUMP SOURCE To start the generator press and hold the ON OFF Switch located in the middle front of the instrument until the GREEN LED turns on approximately 3 seconds Release the switch immediately thereafter The GREEN LED will blink at a rapid rate indicating the number of hours of gas remaining in the source 5 blinks indicates 5 hours of life remaining It will then blink at a rate of once a second for approximately 20 seconds Gas will begin to be generated immediately After 20 seconds the green light will stay illuminated and the pump will begin running With a bump source installed the unit will run for an additional 20 seconds providing a bump of output gas to drive the connected sensor into alarm after which both red and green lights will alternate indicating that the instrument is purging It will shut off automatically after an additional 25 seconds Note Some sensors have faster response times than others If the system does not respond to the gas within 30 seconds continue to leave the MiniBump attached to the sensor throughout the purge cycle and if needed run the bump cycle a second time If the RED LED begins flashing or remains lit continuously it indicates the instrument is in either low battery condition or source failure mode Please refer to the section on Non Normal Operation 7

Page 9

III NORMAL OPERATION MINIBUMP SOURCE The instrument is designed to generate gas for approximately 36 seconds per bump test Each source will provide up to 1000 tests The MiniBump will generate gas until either the battery becomes discharged or the source is consumed If either condition occurs a RED LED will light indicating a non normal condition refer to the Non Normal Operation section that follows The calibration gas is delivered through a small flexible hose attached to the source manifold cap The hose is placed over around or in front of the gas sensor Alternatively the detector can be placed directly in front of the outlet fitting allowing the gas to be presented directly to the detector 8

Page 10

IV START UP MINICAL SOURCE To start the generator press and hold the ON OFF Switch located in the middle front of the instrument until the GREEN LED turns on approximately 3 seconds Release the switch immediately thereafter The GREEN LED will blink at a rapid rate indicating the number of hours of gas remaining in the source 5 blinks indicate 5 hours of life remaining It will then blink at a rate of once a second for approximately 20 seconds Gas will begin to be generated immediately After 20 seconds the pump will come on and the unit will continue running with the green light blinking for an additional 20 seconds After this the green light will stay illuminated indicating the unit is ready to calibrate With a MiniCal source installed the unit will continue to run in this state generating the concentration and type of gas specified by the generating source installed until turned off by pressing the power button for an additional 3 seconds When the unit is turned off it will enter a purge cycle both red and green lights will alternate indicating that the instrument is purging It will shut off automatically after an additional 25 seconds Note Some sensors have faster response times than others Leave the MiniBump calibrator attached to the sensor until the sensor has stabilized and then proceed with the calibration procedure for that sensor If the RED LED begins flashing or remains lit continuously it indicates the instrument is in either low battery condition or source failure mode Please refer to the section on Non Normal Operation 9

Page 11

V NORMAL OPERATION MINICAL SOURCE The instrument is designed to generate calibration gas at 5LPM liters per minute Each MiniCal source will provide 5 hours of calibration gas The MiniBump will generate gas until either the battery becomes discharged or the source is consumed If either condition occurs a RED LED will light indicating a non normal condition refer to the Non Normal Operation section that follows The calibration gas is delivered through a small flexible hose attached to the source manifold cap The hose is placed over around or in front of the gas sensor Alternatively the detector can be placed directly in front of the outlet fitting allowing the gas to be presented directly to the detector When calibrating sample draw instruments the output gas may be collected into a gas collection bag If this method is used it is recommended that the generated gas be allowed to diffuse into a calibration bag for several minutes to partially fill the bag Be sure to disconnect the sample draw unit before the bag is completely emptied of calibration gas 10

Page 12

VI NON NORMAL OPERATION If either the batteries or the generating source become depleted or the generating source is removed from the MiniBump the instrument enters a non normal operating condition and the RED LED will turn on Under a low battery condition the RED LED will begin to flash If the low battery condition occurs during power up the RED LED will flash for approximately 30 seconds and then the instrument will power down If the low battery condition occurs after start up the RED LED will begin flashing and the instrument will continue to run for approximately 30 seconds before power down It is recommended to replace depleted batteries with two AA heavy duty alkaline batteries for maximum operating life If the source becomes depleted or is removed the RED LED will light continuously If the depleted source condition occurs during power up the RED LED will remain on for approximately 30 seconds and then the instrument will power down If the depleted source condition occurs after start up the RED LED will light and the instrument will continue to run for approximately 30 seconds before power down If the source has been depleted it should be removed and replaced with a new generating source Please refer to the section on source life for proper instructions on replacing the generating sources If the source has not been depleted there may be a faulty electrical connection between the source and the instrument Remove the source and inspect the pins protruding from the source and the sockets in the source chamber Remove any debris and or corrosion that may be present and retest the source If either the pins or sockets become damaged or non repairable replace the source chamber and or the source 11

Page 13

VII CELL LIFE The MiniBum p generating source is consumed during normal operation All sources are rated for up to 1000 test maximum 250 minimum Once a source has been depleted it may be replaced with a new source to provide another 1000 tests MiniCal sources last for 5 hours of use Replacement sources may be stored in their containers in a cool humid area for several years under most conditions Different types of generating sources can be used with each instrument The microprocessor in the MiniBump reads the type of gas of the source from a memory chip embedded in the source body It then adjusts itself for each type of gas output For a complete list of correct source part numbers please refer to the parts list to follow To replace the generating source in the instrument first remove the instrument cap located on the top of the instrument The generating source pulls directly up out of the source chamber Do not twist or rotate the generating source while removing it from the source chamber or damage may occur to either the source pins or the source chamber sockets The replacement source is replaced in reverse fashion Care must be taken when replacing the source to correctly line up the electrode pins with the corresponding sockets Do not force the source into the source chamber if it is not properly aligned otherwise the pins may become damaged rendering the source and or instrument useless NOTE Never attach a sample draw instrument directly to the end of the MiniBump delivery cup with no Y or T connector in between the two instruments 12

Page 14

VIII MAINTENANCE The MiniBump instrument is designed to operate at 0 5 liters per minute flow rate For most accurate calibration when using a MiniCal source the flow should be checked and adjusted if necessary Check the flow using a volumetric flow meter connected directly to the output of the instrument If the flow needs to be changed simply adjust the flow control potentiometer until the correct reading is obtained To access the potentiometer simply remove the top cover of the instrument by pulling up on the sides and the two halves will come apart exposing the internal workings of the product Replace the top cover and you are ready to calibrate Changes in altitude and temperature can cause changes in air density thus affecting the output of the MiniCal source The output was calibrated to STP standard temperature and pressure of sea level and 20 C To compensate for differences in altitude or temperature from STP the following tables should be utilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Altitude corrected output in feet multiply by 0 1000 2000 3000 4000 5000 6000 40 50 1 000 1 038 1 076 1 116 1 158 1 202 1 248 1 068 1 102 Temperature Correction Table Temperature corrected output C multiply by 30 20 10 0 10 20 30 0 829 0 863 0 898 0 932 0 966 1 000 1 034 5 PPM 3000 feet and 0 C 5 x 1 116 x 932 5 2 PPM actual output ON OFF Switch The ON OFF Switch is activated through the front membrane panel It is a 13

Page 15

physical switch mounted on the circuit board If the membrane area gets damaged it is replaced as part of the entire front label If the switch becomes nonfunctional it must be repaired or replaced as part of the circuit board Caution Both the electrochemical source and the alkaline batteries used in the MiniBump instrument contain corrosive chemicals While it is not expected that the chemicals in the source or alkaline batteries will leak during normal operation it is recommended that both the source and the batteries be removed from the instrument if it is being stored for periods longer than one week between use Note The generating sources contain very small amounts of acidic solution and should be disposed of as per local or federal regulations For more information please contact the factory 14

Page 16

IX ACCESSORY ITEMS PARTS LIST The following items and spare parts are available for the MiniBump P N Description 115 0400 00 210 0400 00 250 0401 00 400 0500 00 510 0500 01 510 0550 02 510 0570 02 510 0590 02 Screw cap with hose fitting Wall mount case for MiniBump Battery cover MiniGen MiniBump main board Chlorine MiniBump source Hydrogen Sulfide MiniBump source Hydrogen Cyanide MiniBump source Hydrogen MiniBump source 510 0401 0x 510 0405 0x 510 0407 0x 510 0409 0x Chlorine MiniCal source specify PPM 1 25 Hydrogen Sulfide MiniCal source specify PPM 1 25 Hydrogen Cyanide MiniCal source specify PPM 1 25 Hydrogen MiniCal source specify PPM 1 25 715 0405 0X 730 0415 00 910 0550 00 Hose with connector Norprene 5 Hard body water resistant padded accessory case Instruction manual MiniBump 15

Page 17

X STANDARD WARRANTY We warrant gas calibration equipment manufactured and sold by us to be free from defects in materials workmanship and performance for a period of one year from date of shipment Any parts found defective within that period will be repaired or replaced at our option free of charge F O B factory This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service and which must be cleaned repaired or replaced on a routine basis Such items may include a Electrochemical type generating sources b Batteries Warranty is voided by abuse including rough handling mechanical damage alteration or repair procedures not in accordance with the instruction manual This warranty indicates the full extent of our liability and we are not responsible for removal or replacement cost local repair costs transportation costs or contingent expenses incurred without our prior approval Advanced Calibration Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing facilities with transportation charges prepaid and which Advanced Calibration Designs Inc s Material Review Board examination shall disclose to its satisfaction to have been defective This warranty is expressed in lieu of any and all other warranties and representations expressed or implied and all other obligations or liabilities on the part of Advanced Calibration Designs Inc including but not limited to the warranty of fitness for a particular purpose In no event shall Advanced Calibration Designs Inc be liable for direct incidental or consequential loss or damage of any kind connected with the use of its products or failure of its products to function or operate properly The following is a listing of the available electrochemical MiniBump sources and their standard warranty when installed in equipment manufactured and supplied by Advanced Calibration Designs Inc Chlorine Hydrogen Hydrogen Sulfide Hydrogen Cyanide One year or 250 tests 16

Page 18

The following is a listing of the available electrochemical MiniCal sources and their standard warranty when installed in equipment manufactured and supplied by Advanced Calibration Designs Inc Chlorine Hydrogen Hydrogen Sulfide Hydrogen Cyanide One year or 5 hours of calibration gas 17

Page 19

Manufactured by Advanced Calibration Designs Inc 2024 W McMillan Street Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Website www GoACD com 18

Page 20

Advanced Calibration Designs Inc MiniBump Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Telephone 520 290 2855 Fax 520 290 2860 Toll Free 800 737 0223 Email acd goACD com Web www goACD com P N 910 0550 00 Rel IV Calibration Gas Bump Tester Instruction Manual Rel IV